Ecocell 55 - Comfort

  • Convincingly cost saving laboratory dryer
  • 55-liter chamber volume
  • Working temperature of from +5°C over ambient temperature up to 250°C
  • Natural fine air flow in the chamber, noiseless operation
  • High-quality and safe drying, heating and tempering of common laboratory materials in laboratories and industry
  • High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
  • Comfort model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LCD display
  • Wide range of optional equipment and accessories
  • Output for a printer or PC connection
  • Validation possibility (IQ , OQ)


Technical data

Interior of stainless steel material No. 1.4301

  • volume 55 l
  • depth 390 mm
  • width 400 mm
  • height 350 mm

Tray

  • standard equipment 2 pcs
  • storage area (w × d) 380×335 mm
  • racks 4 max. no
  • distance between guide rails 70 mm

Maximal weight of the load *)

  • inside the oven 50 kg/oven
  • per tray 20 kg/tray

Door

  • 1 pcs

External dimensions (including door and handle)

  • depth 640 mm
  • diameter of the air branch outer/inner 52/49 mm
  • width 620 mm
  • height, incl. Feet (N) and Caster (K) 680 (N) mm

Package dimensions (three layers carton)

  • depth 730 mm
  • width 700 mm
  • depth 880 mm

Weight

  • gross 66 kg
  • nett 55 kg

Working temperature (beginning of the regulation)

  • from 5 °C above ambient temp. to: 250 °C

Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door

  • time variation cca (±) 0,3 °C
  • space deviation cca (±) % of the reached temperature 2 %

Time required to reach 250 °C with closed air flap and voltage 230 V

  • 59 min

Heat emission at 250 °C

  • 590 W

Air exchange speed at 250 °C

  • 8 per hour

Electricity - mains 50/60 Hz

  • max. power input 1,2 kW
  • nominal voltage 230 V
  • current 5,2 A
  • input in stand by mode 5 W

*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.

Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.