Friocell 1212 - ECO line

  • Precise, nature-adaptable, cost-saving – laboratory incubator
  • 1212-liter chamber volume
  • Working temperature of from 0°C up to 100°C
  • Patented cooling system (ecological cooling medium R134a) offers a precise and quick simulation of natural processes and reduces the risk of samples evaporation; various types of the inner chamber lighting can be added to the unit
  • High operation comfort; precise temperature regulation; possibility to reach even extremely short times of temperature changes and short times of temperature recovery in the chamber after the door opening; the standard equipment includes inner glass door and possibility to regulate the fan speed
  • Suitable for demanding and accurate tests and samples cultivation
  • Comfort model with a Fuzzy logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LCD display
  • Wide range of optional equipment and accessories
  • Output for a printer or PC connection
  • Validation possibility (IQ , OQ)


Technical data

Inner dimensions - Chamber, stainless steel

  • volume 1212 l
  • depth 520 mm
  • width 3x540 (1905) mm
  • height 1415 mm

Volume of the steam space

  • 1753 cca l

Trays, stainless steel *)

  • Min. distance between trays 70 mm
  • number 3x19/6 max./usual
  • Storage area (w × d) 520 x 485 mm

Number of outer metal doors

  • number 3 pcs

Number of inner glass doors

  • number 3 pcs

Admissible weight of trays

  • together inside the oven 300 kg
  • per 1 tray 30 kg

Electricity data

  • Max. power 2500 W
  • Protective system IP 20
  • mains 50/60Hz 230 V

Temperature data

  • Cooling/down time from 22 °C to 10 °C 21 min
  • Heating/up time to 37 °C from the ambient temperature 30 min
  • Working temperature from 0,0 to 70 °C
  • Temperature accuracy - in time 0,2 ± °C
  • Temperature accuracy - in space at 10 °C 0,6 ± °C

Recovery time after 30 s door open

  • at 37 °C 10 min
  • at 50 °C 10 min

Heat emission

  • at 37 °C 200 W

Outer dimensions, incl. door and handle, Feet (N) or Rollers (K)

  • depth 870 mm
  • width 2435 mm
  • height 1905 (K) mm

Weight

  • brut 839 kg
  • nett 519 kg

Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.

Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.

Contact form

Are you interesting in our offer? Please fill the contact form, we will be pleased to contact you.

Interest in:

Based on the cross marking I hereby express my consent for my data to be collected and saved with the purpose of my inquiry processing. Such consent may be cancelled at any time by sending a message for our attention. You can get detailed information from the Declaration of Data Protection. *

* Required