Vacucell 55 - ECO line

  • Quick, cost-saving, ecological - laboratory oven meeting higher demands
  • 55-liter chamber volume
  • Working temperature of from +5°C over ambient temperature up to 200/300°C
  • Noiseless operation and soft heating of samples in the chamber; high-quality and safe heating and drying of samples to constant mass
  • High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
  • Suitable for perfect drying of thermolabile and oxidation-sensitive materials, complex-shape components, samples drying to constant mass etc. under vacuum
  • Model with a Fuzzy logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • Wide range of optional equipment and accessories (e.g. a “Vacustation” box enabling a selected vacuum pump installation, …)
  • Output for a printer or PC connection
  • Validation possibility (IQ , OQ)


Technical data

Interior of stainless steel material No. 1.4571

  • volume 55 l
  • depth 320 mm
  • width 400 mm
  • height 430 mm

Tray

  • number 2 pcs
  • racks max. 7 pcs

Shelves distance

  • height 43 mm

Usefull dim. of shelf

  • width × depths 340×296 mm

External dimensions (including door and handle)

  • depth 560 mm
  • width 620 mm
  • heights(incl. foots) 910 mm

Package dimensions

  • depth 730 mm
  • width 780 mm
  • height (incl. palette) 1050 mm

Weight

  • brut 111 kg
  • net 98 kg

Working temperature (regulation start)

  • from 5 °C over ambient temp to 200 °C

Temp. deviations acc. to DIN 12 880 Part 2 from working temp – Al shelves pressure 5–10 mbar**

  • in time to: 0,4 ± °C
  • in space, temperature 100 °C do: 2 ± °C
  • in space, temperature 200 °C do: 6 ± °C

Temp. deviations acc. to DIN 12 880 Part 2 from working temp – ss shelves pressure 5–10 mbar**

  • i time to: 1 ± °C
  • in space, temperature 100 °C do: 10 ± °C
  • in space, temperature 200 °C do: 23 ± °C

Time of rise onto 98 % voltage 230 V – (Al shelves pressure 5–10 mbar)

  • onto temp 100 °C 65 min
  • onto temp 200 °C 85 min

Time of rise onto 98 % voltage 230 V – (shelves pressure 5–10 mbar)

  • onto temp 100 °C 140 min
  • onto temp 200 °C 180 min

Heat radiation

  • at 100 °C 260 W
  • at 200 °C 520 W

Vacuum connection

  • feedthrough needle valve for inert gas or air 8 mm
  • measuring acces 40 mm
  • vacuum connection measuring DN 16 mm

Electric parameters – mains 50/60 Hz

  • nominal voltage 230 V
  • mains 50/60 Hz 1,2 kW
  • current 5,2 A
  • input in stand by mode 5 W

Max. permissible load of the shelves

  • inside the oven 45 kg/oven
  • per shelves 25 kg

Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.

Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.